Похожие презентации:
Model 8xl global service training
1. MODEL 8XL GLOBAL SERVICE TRAINING
CAUTION & ERROR CODESCONTINUED
2. GAS EXPOSURE TESTS
Gas Injection Test
Chamber Vacuum Test
Temperature Test
Undefined Errors
3. Kein Folientitel
HEAT SINKTEMP SENSOR
CONTROLLER
VACUUM
SENSOR
H2O
COMPRESSED AIR
4. Kein Folientitel
GAS EXPOSE PHASE55 C
55 o C
750mB
0
12:00
GAS INJECT
E50
STOP
5. Kein Folientitel
FINAL PUMP DOWN55 C
55 o C
750mB
0
12:00
GAS INJECT
E75
STOP
6. Kein Folientitel
GAS EXPOSE PHASE55 C
55 o C
750mB
0
12:00
GAS INJECT
E51
STOP
7. Kein Folientitel
8XL STERILIZERCHAMBER
CHAMBER TEMP SENSORS
HEAT SINK
TEMP SENSOR
CONTROLLER
VACUUM
SENSOR
H2O
COMPRESSED AIR
VACUUM
VENTURI
8. Kein Folientitel
GAS EXPOSE PHASE55 C
55 o C
750mB
0
12:00
GAS INJECT
E52
STOP
9. Kein Folientitel
Gas Expose Phase55 o C
750mB
c3
55 C
0
12:00
PRECONDITION
GAS INJECT
AERATE
STOP
10. Kein Folientitel
FINAL PUMP DOWN55 C
55 o C
750mB
0
12:00
GAS INJECT
E54
STOP
11. Kein Folientitel
FINAL PUMP DOWN55 C
55 o C
750mB
0
12:00
GAS INJECT
E55
STOP
12. Kein Folientitel
FINAL PUMP DOWN55 C
55 o C
750mB
0
12:00
GAS INJECT
E71
STOP
13. Kein Folientitel
FINAL PUMP DOWN55 C
55 o C
750mB
0
12:00
GAS INJECT
E71
STOP
14. Kein Folientitel
FINAL PUMP DOWN55 C
55 o C
750mB
0
12:00
GAS INJECT
E73
STOP
15. Kein Folientitel
FINAL PUMP DOWN55 C
55 o C
750mB
0
12:00
GAS INJECT
STOP
16. Kein Folientitel
FINAL PUMP DOWN55 C
55 o C
750mB
0
12:00
GAS INJECT
E72
STOP
17. Kein Folientitel
Aeration Phase55 C
55 o C
960mB
0
12:00
PRECONDITION
GAS INJECT
AERATE
12:00
2:00
STOP
18. Kein Folientitel
Aeration Phase55 C
55 o C
960mB
0
12:00
AERATE
12:00
E76
2:00
START
STOP
19. Kein Folientitel
Aeration Phase55 C
55 o C
960mB
0
12:00
PRECONDITION
GAS INJECT
AERATE
12:00
12:00
STOP
20. Kein Folientitel
Aeration Phase55 C
55 o C
960mB
0
12:00
AERATE
12:00
E77
2:00
START
STOP
21. Kein Folientitel
8XL STERILIZERCHAMBER
VACUUM
VENTURI
CHAMBER TEMP SENSORS
HEAT SINK
TEMP SENSOR
DOOR LOCK ASSEMBLY
DOOR LATCH
CYLINDER
DOOR LATCH SOL.
CONTROLLER
DOOR LATCH
DOOR LATCH SWITCH
VACUUM
SENSOR
H2O
COMPRESSED AIR
DOOR HANDEL SWTICH
22. AERATION PHASE CAUTION CODES
C3 - POWER INTERRUPTIONC4 - COMPRESSED AIR LOST
C5 - LOSS OF HEATER CONTROL
23. Kein Folientitel
SIDE VIEW POWER LOCATIONSW1
SW2
TEST SW.
24 VDC
SUPPLY
5 VDC
SUPPLY
INTERFACE
CABLES
VENT
PUNCT. SOL.
PUNCTURE CYL.
POWER SWITCH
MAIN POWER
SWITCH
POWER CONNECTOR
WATER SOL.
AIR
24. Kein Folientitel
CHAMBER CHECKVALVE
DOOR LATCH
CYLINDER
PUMP
SOL.
VENTURI
DOOR
SOL.
x
REGULATOR
PRESSURE
GAUGE
AIR
PRESSURE
SWITCH
PUNCTURE CYLINDER
25. Kein Folientitel
8XL STERILIZERCHAMBER
CHAMBER TEMP SENSORS
HEAT SINK
TEMP SENSOR
CONTROLLER
VACUUM
SENSOR
H2O
COMPRESSED AIR
VACUUM
VENTURI
26. Kein Folientitel
Aeration Phase55 C
55 o C
960mB
0
12:00
PRECONDITION
GAS INJECT
AERATE
12:00
12:00
STOP
27. DOOR LOCK ASSEMBLY
DOOR LATCHCYLINDER
DOOR LATCH SOL.
DOOR LATCH
DOOR
SWITCH
DOOR LATCH SWITCH
DOOR HANDLE SWITCH
28. Kein Folientitel
Aeration Phase55 C
55 o C
960mB
0
12:00
PRECONDITION
GAS INJECT
AERATE
12:00
12:00
STOP
29. Kein Folientitel
ERROR CODE CLEARINGGAS EXPOSE PHASE
55 C
55 o C
750mB
0
12:00
GAS INJECT
E51
STOP
30. SILENT, LOCKED DOOR ERROR CODES
E71 - FINAL VACUUM PUMP DOWNE72 - OBSTRUCTED AIR INLET
E73 - VACUUM SENSOR GAS
E75 - GAS LOW PRESSURE
E76 - INTERLOCK RELAY OPEN
E77 - PUNCTURE PIN STUCK
31. ERRORS REQUIRING SERVICE
E73 - PRESSURE SENSOR OUT OF RANGEE75 - GAS CARTRIDGE PARTIALLY FILLED
OR PARTIALLY PUNCTURED.
32. Kein Folientitel
FINAL PUMP DOWN55 C
55 o C
750mB
0
12:00
GAS INJECT
E75
STOP